CVE-2014-5406 Information

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library (2) software update or (3) configuration change which allows remote attackers to modify settings or medication data via packets on the (a) TELNET (b) HTTP (c) HTTPS or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Reference

http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/