CVE-2014-5406 Information
Feb 14, 2021
cve
Description
The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library (2) software update or (3) configuration change which allows remote attackers to modify settings or medication data via packets on the (a) TELNET (b) HTTP (c) HTTPS or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.
Reference
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/ucm446809.htm https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01 https://xs-sniper.com/blog/2015/06/08/hospira-plum-a-infusion-pump-vulnerabilities/