CVE-2014-5433 Information
Feb 14, 2021
cve
Description
An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System Version 8 which incorporates hardware and software changes.
CVSS Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Reference
https://ics-cert.us-cert.gov/advisories/ICSA-15-181-01
Attack Complexity
LOW
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
UNCHANGED
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
HIGH
Base Severity
9.8
Share on: