CVE-2015-3953 Information

Description

Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior Plum A+3 Infusion System version 13.6 and prior and Symbiq Infusion System version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS Vector

CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Reference

https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

Attack Complexity

LOW

Privileges Required

NONE

User Interaction Required

NONE

Scope

NONE

Confidentiality Impact

UNCHANGED

Integrity Impact

HIGH

Availability Impact

HIGH

Base Score

HIGH

Base Severity

9.8

Share on: