CVE-2016-8375 Information
Description
An issue was discovered in Becton Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit Version 9.5 and prior versions and Version 9.7 and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device’s flash memory. The Alaris 8015 PC unit Version 9.7 and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.
CVSS Vector
CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:C/C:H/I:N/A:N
Reference
http://www.securityfocus.com/bid/96113 https://ics-cert.us-cert.gov/advisories/ICSMA-17-017-01 https://ics-cert.us-cert.gov/advisories/ICSMA-17-017-02
Attack Complexity
HIGH
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
CHANGED
Integrity Impact
HIGH
Availability Impact
NONE
Base Score
NONE
Base Severity
4.9
Share on: