A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Version 1.1 1.5 and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however if the wireless association is established and the Ethernet cable is attached the pump does not attach the network stack to the wireless network. In this scenario all network traffic is instead directed over the wired Ethernet connection.

CVSS Vector

CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L

Reference

http://www.securityfocus.com/bid/100665 https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A

Attack Complexity

HIGH

Privileges Required

NONE

User Interaction Required

NONE

Scope

NONE

Confidentiality Impact

UNCHANGED

Integrity Impact

LOW

Availability Impact

LOW

Base Score

LOW

Base Severity

5.6