CVE-2017-5149 Information
Feb 14, 2021
cve
Description
An issue was discovered in St. Jude Medical Merlin@home versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical’s web site Merlin.net are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.
CVSS Vector
CVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
Reference
http://www.securityfocus.com/bid/95331 https://ics-cert.us-cert.gov/advisories/ICSMA-17-009-01A
Attack Complexity
HIGH
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
CHANGED
Integrity Impact
LOW
Availability Impact
HIGH
Base Score
HIGH
Base Severity
8.9