CVE-2018-10631 Information
Feb 14, 2021
cve
Description
Medtronic N’Vision Clinician Programmer 8840 N’Vision Clinician Programmer all versions and 8870 N’Vision removable Application Card all versions. The 8840 Clinician Programmer executes the application program from the 8870 Application Card. An attacker with physical access to an 8870 Application Card and sufficient technical capability can modify the contents of this card including the binary executables. If modified to bypass protection mechanisms this malicious code will be run when the card is inserted into an 8840 Clinician Programmer.
CVSS Vector
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Reference
https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01 https://www.medtronic.com/security
Attack Complexity
LOW
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
UNCHANGED
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
HIGH
Base Severity
6.8