CVE-2019-18254 Information

Description

BIOTRONIK CardioMessenger II The affected products do not encrypt sensitive information while at rest. An attacker with physical access to the CardioMessenger can disclose medical measurement data and the serial number from the implanted cardiac device the CardioMessenger is paired with.

CVSS Vector

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Reference

https://www.us-cert.gov/ics/advisories/icsma-20-170-05

Attack Complexity

LOW

Privileges Required

NONE

User Interaction Required

NONE

Scope

NONE

Confidentiality Impact

UNCHANGED

Integrity Impact

HIGH

Availability Impact

NONE

Base Score

NONE

Base Severity

4.6