Description

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

CVSS Vector

CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Reference

https://talosintelligence.com/vulnerability_reports/TALOS-2019-0785

Attack Complexity

LOW

Privileges Required

NONE

User Interaction Required

NONE

Scope

NONE

Confidentiality Impact

CHANGED

Integrity Impact

HIGH

Availability Impact

HIGH

Base Score

HIGH

Base Severity

7.6