CVE-2020-12048 Information

Description

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40 The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g. TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

CVSS Vector

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Reference

https://www.us-cert.gov/ics/advisories/icsma-20-170-03

Attack Complexity

LOW

Privileges Required

NONE

User Interaction Required

NONE

Scope

NONE

Confidentiality Impact

UNCHANGED

Integrity Impact

HIGH

Availability Impact

NONE

Base Score

NONE

Base Severity

7.5