CVE-2020-15486 Information
Feb 14, 2021
cve
Description
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also saved data can also be extracted over a Bluetooth connection. In addition an attacker can launch a man-in-the-middle attack against data integrity.
CVSS Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
Reference
Attack Complexity
LOW
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
UNCHANGED
Integrity Impact
HIGH
Availability Impact
NONE
Base Score
NONE
Base Severity
6.5
Share on: