CVE-2020-7590 Information
Feb 14, 2021
cve
Description
A vulnerability has been identified in DCA Vantage Analyzer (All versions V4.5 are affected by CVE-2020-7590. In addition serial numbers 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device.
CVSS Vector
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Reference
https://www.siemens-healthineers.com/support-documentation/security-advisory
Attack Complexity
LOW
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
UNCHANGED
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
HIGH
Base Severity
6.8
Share on: