CVE-2021-37160 Information
Description
A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g. cryptographic signature validation) during a File Upload for a firmware update.
CVSS Vector
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Reference
https://www.swisslog-healthcare.com/en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%20%20CVE-2021-37164%20%204%20more%20rows%20 https://www.armis.com/PwnedPiper https://www.swisslog-healthcare.com https://www.swisslog-healthcare.com/-/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37160-bulletin—no-firmware-update-validation.pdf?rev=c7f94647037c4007992e2e626d445561&hash=E89531490070A809FB74994018BA1248
Attack Complexity
LOW
Privileges Required
NONE
User Interaction Required
NONE
Scope
NONE
Confidentiality Impact
UNCHANGED
Integrity Impact
HIGH
Availability Impact
HIGH
Base Score
HIGH
Base Severity
9.8
Share on: