There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.